{ "containers": { "cna": { "affected": [ { "product": "LifeCare PCA Infusion System", "vendor": "Hospira", "versions": [ { "status": "affected", "version": "<= 5.0" } ] } ], "datePublic": "2015-05-13T00:00:00.000Z", "descriptions": [ { "lang": "en", "value": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access." } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-312", "description": "Cleartext storage of sensitive information CWE-312", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2019-03-25T18:20:12.000Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01" } ], "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2015-1012", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "LifeCare PCA Infusion System", "version": { "version_data": [ { "version_value": "<= 5.0" } ] } } ] }, "vendor_name": "Hospira" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "Cleartext storage of sensitive information CWE-312" } ] } ] }, "references": { "reference_data": [ { "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01" } ] } } }, "adp": [ { "providerMetadata": { "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE", "dateUpdated": "2024-08-06T04:26:11.620Z" }, "title": "CVE Program Container", "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01" } ] } ] }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2015-1012", "datePublished": "2019-03-25T18:20:12.000Z", "dateReserved": "2015-01-10T00:00:00.000Z", "dateUpdated": "2024-08-06T04:26:11.620Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }