{ "containers": { "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "24950 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "24952 MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Peter Morgan of Clever Security reported this vulnerability" } ], "datePublic": "2018-06-29T06:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "
\nMedtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.\n\n
" } ], "value": "Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "HIGH", "attackVector": "PHYSICAL", "availabilityImpact": "LOW", "baseScore": 6.2, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "LOW", "privilegesRequired": "LOW", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-749", "description": "CWE-749", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert", "dateUpdated": "2025-05-22T18:14:07.710Z" }, "references": [ { "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-18.html" }, { "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01" } ], "solutions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "Medtronic has released additional patient focused information, at the following location:
https://www.medtronic.com/security
\n\nMedtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Medtronic has released additional patient focused information, at the following location:
https://www.medtronic.com/security
\n\n