{ "dataType": "CVE_RECORD", "dataVersion": "5.2", "cveMetadata": { "cveId": "CVE-2019-25720", "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10", "state": "PUBLISHED", "assignerShortName": "VulnCheck", "dateReserved": "2026-06-02T14:22:01.500Z", "datePublished": "2026-06-03T16:56:29.683Z", "dateUpdated": "2026-06-03T17:44:47.456Z" }, "containers": { "cna": { "providerMetadata": { "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10", "shortName": "VulnCheck", "dateUpdated": "2026-06-03T16:56:29.683Z" }, "title": "Dräger SC Monitoring Devices DoS via Malformed Network Packet", "datePublic": "2019-11-21T00:00:00.000Z", "problemTypes": [ { "descriptions": [ { "lang": "en", "cweId": "CWE-1286", "description": "CWE-1286 Improper Validation of Syntactic Correctness of Input", "type": "CWE" } ] } ], "affected": [ { "vendor": "Dräger", "product": "SC 6002XL", "versions": [ { "status": "affected", "version": "SC 6002XL", "versionType": "custom" } ], "defaultStatus": "unknown" }, { "vendor": "Dräger", "product": "SC6802XL", "versions": [ { "status": "affected", "version": "SC6802XL", "versionType": "custom" } ], "defaultStatus": "unknown" }, { "vendor": "Dräger", "product": "SC 7000", "versions": [ { "status": "affected", "version": "SC 7000", "versionType": "custom" } ], "defaultStatus": "unknown" }, { "vendor": "Dräger", "product": "SC8000", "versions": [ { "status": "affected", "version": "SC8000", "versionType": "custom" } ], "defaultStatus": "unknown" }, { "vendor": "Dräger", "product": "SC90000 XL", "versions": [ { "status": "affected", "version": "SC90000 XL", "versionType": "custom" } ], "defaultStatus": "unknown" } ], "descriptions": [ { "lang": "en", "value": "Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.", "supportingMedia": [ { "type": "text/html", "base64": false, "value": "
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
" } ] } ], "references": [ { "url": "https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf", "tags": [ "vendor-advisory" ] }, { "url": "https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet", "tags": [ "third-party-advisory" ] } ], "metrics": [ { "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ], "cvssV4_0": { "attackVector": "ADJACENT", "attackComplexity": "LOW", "attackRequirements": "NONE", "privilegesRequired": "NONE", "userInteraction": "NONE", "vulnConfidentialityImpact": "NONE", "subConfidentialityImpact": "NONE", "vulnIntegrityImpact": "NONE", "subIntegrityImpact": "NONE", "vulnAvailabilityImpact": "HIGH", "subAvailabilityImpact": "NONE", "exploitMaturity": "NOT_DEFINED", "Safety": "NOT_DEFINED", "Automatable": "NOT_DEFINED", "Recovery": "NOT_DEFINED", "valueDensity": "NOT_DEFINED", "vulnerabilityResponseEffort": "NOT_DEFINED", "providerUrgency": "NOT_DEFINED", "version": "4.0", "baseSeverity": "HIGH", "baseScore": 7.1, "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N" } }, { "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ], "cvssV3_1": { "version": "3.1", "attackVector": "ADJACENT_NETWORK", "attackComplexity": "LOW", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "UNCHANGED", "confidentialityImpact": "NONE", "integrityImpact": "NONE", "availabilityImpact": "HIGH", "baseSeverity": "MEDIUM", "baseScore": 6.5, "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H" } } ], "credits": [ { "lang": "en", "value": "Jeroen Slobbe and Max Grim", "type": "finder" } ], "source": { "discovery": "EXTERNAL" } }, "adp": [ { "metrics": [ { "other": { "type": "ssvc", "content": { "timestamp": "2026-06-03T17:44:08.218138Z", "id": "CVE-2019-25720", "options": [ { "Exploitation": "none" }, { "Automatable": "no" }, { "Technical Impact": "partial" } ], "role": "CISA Coordinator", "version": "2.0.3" } } } ], "title": "CISA ADP Vulnrichment", "providerMetadata": { "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP", "dateUpdated": "2026-06-03T17:44:47.456Z" } } ] } }