{ "containers": { "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "SureSigns VS4", "vendor": "Philips", "versions": [ { "lessThan": "A.07.107", "status": "affected", "version": "0", "versionType": "custom" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Cleveland Clinic reported these vulnerabilities to Philips." } ], "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "

Philips SureSigns VS4, A.07.107 and prior receives input or data, but it does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly.

" } ], "value": "Philips SureSigns VS4, A.07.107 and prior receives input or data, but it does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "LOW", "baseScore": 2.1, "baseSeverity": "LOW", "confidentialityImpact": "NONE", "integrityImpact": "NONE", "privilegesRequired": "LOW", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:N/I:N/A:L", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-20", "description": "CWE-20", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert", "dateUpdated": "2025-06-04T21:32:44.371Z" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-233-01" }, { "url": "https://www.philips.com/a-w/security/security-advisories/product-security-2020.html#2020_archive" } ], "source": { "advisory": "ICSMA-20-233-01", "discovery": "EXTERNAL" }, "title": "Philips SureSigns VS4 Improper Input Validation", "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "As a mitigation to these vulnerabilities, Philips recommends users \nchange all system passwords on the SureSigns VS4 with unique passwords \nfor each device and secure the device when not in use to prevent \nunauthorized access, as referenced in the Installation and Configuration\n Guide available on Incenter. Philips also recommends users consider \nreplacing the SureSigns VS4 device with a newer technology.

\nUsers with questions regarding specific SureSigns VS4 patient monitor installations and upgrade options should contact Philips service support or regional service support or call 1-800-722-9377.

\nPlease see the Philips advisory for vulnerabilities discussed in this disclosure, and visit the Philips product security website for the latest security information for Philips products.\n\n
" } ], "value": "As a mitigation to these vulnerabilities, Philips recommends users \nchange all system passwords on the SureSigns VS4 with unique passwords \nfor each device and secure the device when not in use to prevent \nunauthorized access, as referenced in the Installation and Configuration\n Guide available on Incenter. Philips also recommends users consider \nreplacing the SureSigns VS4 device with a newer technology.\n\n\nUsers with questions regarding specific SureSigns VS4 patient monitor installations and upgrade options should contact Philips service support or regional service support https://www.usa.philips.com/healthcare/solutions/customer-service-solutions or call 1-800-722-9377.\n\n\nPlease see the Philips advisory http://www.philips.com/productsecurity for vulnerabilities discussed in this disclosure, and visit the Philips product security website https://www.philips.com/productsecurity for the latest security information for Philips products." } ], "x_generator": { "engine": "Vulnogram 0.2.0" }, "x_legacyV4Record": { "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-16237", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "Philips SureSigns VS4", "version": { "version_data": [ { "version_value": "A.07.107 and prior" } ] } } ] }, "vendor_name": "n/a" } ] } }, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Philips SureSigns VS4, A.07.107 and prior. The product receives input or data, but it does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly." } ] }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "IMPROPER INPUT VALIDATION CWE-20" } ] } ] }, "references": { "reference_data": [ { "name": "https://us-cert.cisa.gov/ics/advisories/icsma-20-233-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-233-01" } ] } } }, "adp": [ { "providerMetadata": { "orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE", "dateUpdated": "2024-08-04T13:37:54.192Z" }, "title": "CVE Program Container", "references": [ { "tags": [ "x_refsource_MISC", "x_transferred" ], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-233-01" } ] } ] }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-16237", "datePublished": "2020-08-21T12:11:41.000Z", "dateReserved": "2020-07-31T00:00:00.000Z", "dateUpdated": "2025-06-04T21:32:44.371Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1" }