{ "dataType": "CVE_RECORD", "dataVersion": "5.1", "cveMetadata": { "cveId": "CVE-2024-48967", "assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa", "state": "PUBLISHED", "assignerShortName": "Baxter", "dateReserved": "2024-10-10T19:24:34.436Z", "datePublished": "2024-11-14T21:40:44.904Z", "dateUpdated": "2024-11-15T14:28:13.605Z" }, "containers": { "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "Life2000 Ventilation System", "vendor": "Baxter", "versions": [ { "status": "affected", "version": "06.08.00.00 and prior" } ] } ], "datePublic": "2024-11-14T21:38:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance." } ], "value": "The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance." } ], "impacts": [ { "capecId": "CAPEC-117", "descriptions": [ { "lang": "en", "value": "CAPEC-117 Interception" } ] }, { "capecId": "CAPEC-441", "descriptions": [ { "lang": "en", "value": "CAPEC-441 Malicious Logic Insertion" } ] } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "NETWORK", "availabilityImpact": "HIGH", "baseScore": 10, "baseSeverity": "CRITICAL", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-778", "description": "CWE-778: Insufficient Logging", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "orgId": "dba971b9-eb30-4121-91e1-3b45611354aa", "shortName": "Baxter", "dateUpdated": "2024-11-14T21:52:23.702Z" }, "references": [ { "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01" } ], "source": { "discovery": "UNKNOWN" }, "title": "Life2000 ventilator and Service PC lack sufficient audit logging capabilities", "x_generator": { "engine": "Vulnogram 0.2.0" } }, "adp": [ { "affected": [ { "vendor": "baxter", "product": "life2000_ventilator_firmware", "cpes": [ "cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*" ], "defaultStatus": "unknown", "versions": [ { "version": "0", "status": "affected", "lessThan": "6.08.00.00", "versionType": "custom" } ] } ], "metrics": [ { "other": { "type": "ssvc", "content": { "timestamp": "2024-11-15T14:28:10.676566Z", "id": "CVE-2024-48967", "options": [ { "Exploitation": "none" }, { "Automatable": "yes" }, { "Technical Impact": "total" } ], "role": "CISA Coordinator", "version": "2.0.3" } } } ], "title": "CISA ADP Vulnrichment", "providerMetadata": { "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP", "dateUpdated": "2024-11-15T14:28:13.605Z" } } ] } }