METADATA
last updated: 2025-12-14 RT
file_name: SOP-105-A_1 QuickColor Test Diagrams.md
file_date: 2022-08-26
title: FloodLAMP QMS SOP-105-A_1 QuickColor Test Diagrams
category: guides
subcategory: qms-sops
tags:
source_file_type: gdoc
xfile_type: docx
gfile_url: https://docs.google.com/document/d/15oCAL2T0or_BigVAw1kpXS_BojPpcPsMOw8aJ_qw5mA
xfile_github_download_url: https://raw.githubusercontent.com/FocusOnFoundationsNonprofit/floodlamp-archive/main/guides/qms-sops/SOP-105-A_1%20QuickColor%20Test%20Diagrams.docx
pdf_gdrive_url: https://drive.google.com/file/d/1hgxWFnT6RCJRAA6CyJWD1lnAr_5P1vNT
pdf_github_url: https://github.com/FocusOnFoundationsNonprofit/floodlamp-archive/blob/main/guides/qms-sops/SOP-105-A_1%20QuickColor%20Test%20Diagrams.pdf
conversion_input_file_type: docx
conversion: pandoc
license: CC BY 4.0 - https://creativecommons.org/licenses/by/4.0/
tokens: 632
words: 304
notes:
summary_short: The FloodLAMP QMS SOP-105-A_1 QuickColor Test Diagrams provides step-by-step diagrams for running the FloodLAMP QuickColor COVID-19 pooled saliva test, covering inactivation solution preparation, reaction mix formulation, sample processing, and colorimetric amplification readout. It serves as a visual bench reference for lab technicians performing the assay.
CONTENT
FloodLAMP Biotechnologies
**Document Number:** SOP-105-A_1
**Effective Date:** 08/26/2022
**Version:** 1
***INTERNAL TITLE:*** QuickColor Test Diagrams
Print Format
## FloodLAMP QuickColor COVID-19 Test Flow
_Description based on Diagram_
1\) **Inactivation solution (prepare)**
- Combine:
- Saline
- 100X Inactivation Solution
- Apply to: Pooled dry swab samples
- Heat treatment:
- 95 °C
- 5 mL: 8 min
- 1.5 mL: 5 min
- Cool down:
- Let cool
- 5 mL: 10 min
- 1.5 mL: 5 min
2\) **Reaction mix (prepare)**
- Combine:
- Primer solution
- Color LAMP Master Mix
3\) **Run reaction**
- Add processed samples to reaction mix
- Incubate:
- 65 °C for 25 min
## FloodLAMP QuickColor COVID-19 Test Diagram
_Description based on Diagram_
1\) Make 1X Inactivation Saline Solution
- **Add 100X Inactivation Solution to 0.9% Saline**
| | |
|:----------------------------:|:------------------------:|
| **100X Isol. Saline**
**500 µL** | **0.9% Saline**
**50 mL** |
||
_Result: 1X Inactivation Saline Solution (~50 mL)_
---
2\) Elute Pooled Dry Swab Samples
- **Add 1,000 µL of 1X Inactivation Saline Solution to pooled dry swab samples**
_Result: Eluted sample (1,000 µL)_
---
3\) Heat Inactivate & Cool
- **Heat at 95°C for 8 min**
- **Cool at room temperature for 10 min**
---
4\) Make Reaction Mix (CLAMP Reaction Mix)
- **In a 1.5 mL tube, combine:**
| | | | | |
|:----------:|:---------:|:----------:|:----------:|:----------:|
| **Reactions** | **8** | **16** | **24** | **48** |
| **PGS (µL)** | 92 | 183 | 275 | 550 |
| **LAMP MM (µL)** | 109 | 218 | 328 | 655 |
||
---
5\) Set Up Reactions
- **Dispense 23 µL reaction mix per tube (well)**
- **Add 2 µL sample**
- **Include Positive and Negative controls**
---
6\) Amplify & Read
- **Heat at 65°C for 25 min**
- _Output: Results (colorimetric readout)_