Validation Log 2024-12-01 06:01:44.180019 mwtab Python Library Version: 1.2.5 Source: https://www.metabolomicsworkbench.org/rest/study/analysis_id/AN004514/mwtab/txt Study ID: ST002773 Analysis ID: AN004514 File format: txt Status: Contains Validation Errors Number Errors: 1 Error Log: SCHEMA: Section "CHROMATOGRAPHY" does not match the allowed schema. Wrong key 'SOLVENT_C' in OrderedDict([('CHROMATOGRAPHY_SUMMARY', 'The C18 column is operated parallel to the HILIC column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6-port switching valves. During operation of the C18 method, the MS is operated in negative ion mode and 10 μL of sample is injected onto the C18 column while the HILIC column is flushing with wash solution. Flow rate is maintained at 0.4 mL/min until 1.5 min, increased to 0.5 mL/min at 2 min and held for 3 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 10mM ammonium acetate in LC-MS grade water. Initial mobile phase conditions are 60% A, 35% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3.5 min, resulting in a total analytical run time of 5 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 2.5 min, followed by an equilibration solution of 60% A, 35% B, 5% C for 2.5 min.'), ('CHROMATOGRAPHY_TYPE', 'Reversed phase'), ('INSTRUMENT_NAME', 'Thermo Dionex Ultimate 3000 RS'), ('COLUMN_NAME', 'Waters XBridge BEH Amide (50 x 2.1mm,2.5um)'), ('SOLVENT_A', '100% water'), ('SOLVENT_B', '100% acetonitrile'), ('SOLVENT_C', '100% water; 10 mM ammonium acetate'), ('FLOW_GRADIENT', 'Initial mobile phase conditions are 60% A, 35% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3.5 min, resulting in a total analytical run time of 5 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 2.5 min, followed by an equilibration solution of 60% A, 35% B, 5% C for 2.5 min.'), ('FLOW_RATE', '0.4-0.5 mL/min'), ('COLUMN_TEMPERATURE', '40'), ('METHODS_ID', 'EmoryUniversity_HRM_DC5min_082016_01.pdf')])