{ "document": { "acknowledgments": [ { "organization": "Medtronic", "summary": "reporting this vulnerability to CISA" } ], "category": "csaf_security_advisory", "csaf_version": "2.0", "distribution": { "text": "Disclosure is not limited", "tlp": { "label": "WHITE", "url": "https://us-cert.cisa.gov/tlp/" } }, "lang": "en-US", "notes": [ { "category": "general", "text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov", "title": "CISA Disclaimer" }, { "category": "legal_disclaimer", "text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.", "title": "Legal Notice" }, { "category": "summary", "text": "Successful exploitation of this vulnerability could cause the clinician application’s custom password to be reset to default, resulting in unauthorized control of the clinician therapy application.", "title": "Risk evaluation" }, { "category": "other", "text": "Healthcare and Public Health", "title": "Critical infrastructure sectors" }, { "category": "other", "text": "Worldwide", "title": "Countries/areas deployed" }, { "category": "other", "text": "Ireland", "title": "Company headquarters location" }, { "category": "general", "text": "CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.", "title": "Recommended Practices" }, { "category": "general", "text": "CISA also provides a section for control systems security recommended practices on the ICS webpage at cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.", "title": "Recommended Practices" }, { "category": "general", "text": "Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.", "title": "Recommended Practices" }, { "category": "general", "text": "Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.", "title": "Recommended Practices" }, { "category": "general", "text": "CISA also recommends users take the following measures to protect themselves from social engineering attacks:", "title": "Recommended Practices" }, { "category": "general", "text": "Do not click web links or open attachments in unsolicited email messages.", "title": "Recommended Practices" }, { "category": "general", "text": "Refer to Recognizing and Avoiding Email Scams for more information on avoiding email scams.", "title": "Recommended Practices" }, { "category": "general", "text": "Refer to Avoiding Social Engineering and Phishing Attacks for more information on social engineering attacks.", "title": "Recommended Practices" }, { "category": "general", "text": "No known public exploits specifically target this vulnerability. This vulnerability is not exploitable remotely.", "title": "Recommended Practices" } ], "publisher": { "category": "coordinator", "contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870", "name": "CISA", "namespace": "https://www.cisa.gov/" }, "references": [ { "category": "self", "summary": "ICS Advisory ICSMA-23-061-01 JSON", "url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2023/icsma-23-061-01.json" }, { "category": "self", "summary": "ICS Advisory ICSMA-23-061-01 Web Version", "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-061-01" }, { "category": "external", "summary": "Recommended Practices", "url": "https://us-cert.cisa.gov/ics/Recommended-Practices" }, { "category": "external", "summary": "Recommended Practices", "url": "https://cisa.gov/ics" }, { "category": "external", "summary": "Recommended Practices", "url": "https://www.cisa.gov/uscert/sites/default/files/publications/emailscams0905.pdf" }, { "category": "external", "summary": "Recommended Practices", "url": "https://www.cisa.gov/uscert/ncas/tips/ST04-014" } ], "title": "Medtronic Micro Clinician and InterStim Apps", "tracking": { "current_release_date": "2023-03-06T16:35:54.956247Z", "generator": { "engine": { "name": "CISA CSAF Generator", "version": "1.0.0" } }, "id": "ICSMA-23-061-01", "initial_release_date": "2023-03-06T16:35:54.956247Z", "revision_history": [ { "date": "2023-03-06T16:35:54.956247Z", "legacy_version": "1", "number": "1", "summary": "CSAF Creation Date" } ], "status": "final", "version": "1" } }, "product_tree": { "branches": [ { "branches": [ { "branches": [ { "category": "product_version", "name": "A51200", "product": { "name": "Micro Clinician: A51200", "product_id": "CSAFPID-0001" } } ], "category": "product_name", "name": "Micro Clinician" }, { "branches": [ { "category": "product_version", "name": "A51300", "product": { "name": "InterStim X Clinician: A51300", "product_id": "CSAFPID-0002" } } ], "category": "product_name", "name": "InterStim X Clinician" } ], "category": "vendor", "name": "Medtronic " } ] }, "vulnerabilities": [ { "cve": "CVE-2023-25931", "cwe": { "id": "CWE-620", "name": "Unverified Password Change" }, "notes": [ { "category": "summary", "text": "Medtronic Clinician (A51200) and InterStim X Clinicain App (A51300) contain a vulnerability that exists under certain reset conditions, which could cause the clinician application’s custom password to be reset to a default password. This could result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. To gain unauthorized access, an individual would need physical access to the Smart Programmer. [CVE-2023-25931(http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2023-25931) has been assigned to this vulnerability. A CVSS v3 base score of 6.4 has been assigned; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H).", "title": "Vulnerability Summary" } ], "product_status": { "known_affected": [ "CSAFPID-0001", "CSAFPID-0002" ] }, "references": [ { "category": "external", "summary": "web.nvd.nist.gov", "url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2023-25931" }, { "category": "external", "summary": "www.first.org", "url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H" } ], "remediations": [ { "category": "vendor_fix", "details": "The following mitigations have been provided by Medtronic:", "product_ids": [ "CSAFPID-0001", "CSAFPID-0002" ] }, { "category": "vendor_fix", "details": "An app update is available as of February 23, 2023 that will fix the vulnerability.", "product_ids": [ "CSAFPID-0001", "CSAFPID-0002" ] }, { "category": "vendor_fix", "details": "Users should refer to the Medtronic Security Bulletin for the correct Medtronic Support contact for help updating the app.", "product_ids": [ "CSAFPID-0001", "CSAFPID-0002" ], "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/pelvic-health-interstim-micro.html" } ], "scores": [ { "cvss_v3": { "baseScore": 6.4, "baseSeverity": "MEDIUM", "vectorString": "CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H", "version": "3.0" }, "products": [ "CSAFPID-0001", "CSAFPID-0002" ] } ] } ] }